Rigol DS1000E Performance check in Peak Detect mode (Windows mode)
Loading...
3,504
Loading...
Uploader Comments (InstrumentVu)
see all
All Comments (5)
-
Hi, Change the memory depth to "Long Mem" and this appears to cure the problem?
-
I buy DS1102E is not good. . 1. set SINGLE Trigger -> capture pulse -> H.Scale, ZoomIn ZoomIn... keep Square, yes! is nice.. . 2. set NORMAL Trigger -> capture pulse -> H.Scale, ZoomIn ZoomIn... triangle wave!, oh! is very bad .. . . my company buy GDS-1152A is good! . 1. set SINGLE Trigger -> capture pulse -> H.Scale, ZoomIn ZoomIn.... keep Square, yes! is nice.. . 2. set NORMAL Trigger -> capture pulse -> H.Scale, ZoomIn ZoomIn.... keep Square, yes! is nice too..
-
Hi. I don't have that scope, but I think I know what it's doing.
It's probably not an acquisition error.
The graphics routines simply connect min and max point of each sampling interval in succession: min1,max1,min2,max2,... in order to fill the area in between.
This works when points are horizontally less than one pixel apart (upper trace), otherwise you see diagonal lines connecting them (bottom trace).
I use this drawing method in one of my own programs.
-
Hi, I tried the same settings with mine and I get the same response.
S/N: DS1ED110300375
FW: 00.01.06
Why would you say it is a problem?
Loading...
I have one DS1052E I don't know what you want to know.
Twistx77 2 years ago
Hi~! Sorry for delay! May be you can help me to check your equipment have a same issue too. If it is true, please tell me youe SN number and FW version, I will try to transfer this problem to that company, thanks for your help.
InstrumentVu 2 years ago