Smiths Medical is recalling certain Portex uncuffed tracheal tubes in pediatric sizes. The recall affects several models of 2.5, 3.0 and 3.5 mm tubes manufactured before September 2009.
The internal diameter of some of these tracheal tubes is slightly smaller than stated on the labeling. This could make it difficult to insert or withdraw a suction catheter to remove secretions, which could lead to airway obstruction and an inability to ventilate the patient. The defect could also increase airway resistance, and this too could compromise the ability to ventilate the patient. In September, the company notified customers to return all unused units of the recalled tubes and provided recommendations for managing products that are currently in use.