It is failure to sense and failure to capture in the same time ,there is not capture beat after the spike before the sensing beat
may be because the setting is inappropriate in both of pacing output or pathology of the heart and low sensitivity setting
aiwphich 11 months ago
Looks more like VOO pacing with intermittent non-capture. As in the beginne there IS capture for the first 3 beats. The 4th spikes has no QRS so: non-capture. In VOO modus there is no sensing!
aardvarkje 2 years ago
It is failure to sense and failure to capture in the same time ,there is not capture beat after the spike before the sensing beat
may be because the setting is inappropriate in both of pacing output or pathology of the heart and low sensitivity setting
aiwphich 11 months ago
Looks more like VOO pacing with intermittent non-capture. As in the beginne there IS capture for the first 3 beats. The 4th spikes has no QRS so: non-capture. In VOO modus there is no sensing!